日本質量分析学会 第70回質量分析総合討論会会

演題概要

ポスター発表

第1日 6月22日(水)  P会場(501,502,503)

Matrixフリーレーザー脱離イオン化法:DIUTHAMEの高感度化

(浜松ホトニクス)
o小谷政弘大村孝幸井上圭祐池田貴将

We developed DIUTHAME (Desorption Ionization Using Through Hole Alumina MEmbrane) which enables measurements of MALDI without matrix compounds. Output signal from DIUTHAME is sometimes lower than that of MALDI, partly because of its low absorption of laser energy. This study aims to improve the absorption efficiency of laser energy in order to increase the signal intensity of DIUTHAME. As improvement method, the through-holes of DIUTHAME were developed into a tapered shape, and metal nanoparticles were formed on the Pt conductive layer of DIUTHAME surface. From the results of MS imaging of mouse brain sections, it was confirmed that the average signal intensity of phospholipids was improved by 10 times.