ポスター発表
- 第3日 5月13日(水) P会場(1008/09)
-
3P-10 PDF
ポリマー系ジオール基型HILICカラムを用いたオリゴ核酸のLC/MS測定
For development or quality control of oligonucleotide medicines, it is required to analyze them with high sensitivity and high selectivity using LC/MS. Though ion-pair reverse-phase chromatography have been used conventionally, there is a problem that ion-pair reagent tend to remain in LC/MS instrument. In order to solve this issue, a HILIC column Shodex® HILICpak® VN-50 2D with polyvinyl alcohol base packing material modified with diol group was used for the LC. Aqueous solution including ammonium formate / acetonitrile was used as eluent and ESI-MS was used for detection. As a result, it was confirmed to obtain good separation of 2-20mer DNA. This method enables the analysis of oligonucleotides without ion-pair reagent.