ポスター発表
第2日 6月11日(火) P会場(多目的ホール・大会議室101+102)P1会場(多目的ホール)・P2会場(会議室101+102)
- 2P-12
t-SPESI-質量分析イメージングによる高分子フィルム用光安定剤の光劣化反応生成物の可視化
(1阪大院理・ 2東レリサーチセ・ 3慶大先端生命研)
o秋山毅1,2・ 大塚洋一1・ 孫夢沢1・ 平山明由3・ 豊田岐聡1
Mass spectrometry imaging of a polyethylene film containing Chimassorb® 944 as the HALS (hindered amine light stabilizer) with photodegraded areas was performed using t-SPESI (tapping-mode scanning probe electrospray ionization) after optimizing the extraction solvent with LESA (liquid extraction surface analysis). We found that the optimal extraction solvents differ between LESA and t-SPESI. We discovered that Chimassorb® 944 was localized in the non-irradiated area compared to the light-irradiated area, and its oxidative decomposition product and the low molecular weight component were localized in the light-irradiated area and intermediate area.