LC/MS analysis of oligonucleotides using polymer-based HILIC column having diol group

(Showa Denko)
oEiji Kagawa, Junji Sasuga

For development or quality control of oligonucleotide medicines, it is required to analyze them with high sensitivity and high selectivity using LC/MS. Though ion-pair reverse-phase chromatography have been used conventionally, there is a problem that ion-pair reagent tend to remain in LC/MS instrument. In order to solve this issue, a HILIC column Shodex® HILICpak® VN-50 2D with polyvinyl alcohol base packing material modified with diol group was used for the LC. Aqueous solution including ammonium formate / acetonitrile was used as eluent and ESI-MS was used for detection. As a result, it was confirmed to obtain good separation of 2-20mer DNA. This method enables the analysis of oligonucleotides without ion-pair reagent.