LC-MS/MSを用いた血中濃度測定に可溶化剤が与える影響

(小野薬品)
o加藤純也松浦有理山浦由之今若治夫

At the early stages of drug development, the solubilizing agents are used to dissolve poorly soluble compounds with dosing solution. Depending on the route of administration, there is a possibility that co-eluting solubilizing agents affect the ionization of the target compounds in LC-MS/MS analysis. Ion suppression of the target compounds was determined in the plasma with addition of various solubilizing agents. Polyethylene glycol 400, tween 80 and solutol HS15 affected the ionization of the target compounds. The result suggested the possibility of underestimating the actual plasma concentration by some solubilizing agents.