LC/MS用分離技術「真空示差イオンモビリティ分析装置」の開発

(1島津2島津(欧州)3ジュネーブ大)
o洪暎淳1石井俊匡1矢嶋敦1前田裕樹1小河潔1古橋治1Knight, Patrick2Entwistle, Andrew2Girard, Maria Fernanda Cifuentes3Hopfgartner, Gerard3

Differential mobility spectrometry (DMS) separates ions based on the nonlinear change in their mobility between high to low electric fields, which are elicited in a field asymmetric waveform applied to electrodes. This nonlinear change in mobility has less dependence on a molecular weight, and thus DMS can be useful for separation of structural isomers and isobars. The mobility coefficient in DMS can be expanded by nonlinear terms of E/N (E: Electric field strength, N: Gas density). Therefore, higher E/N gives greater change in mobility and result in higher resolution. We have developed vacuum DMS, which reduces gas density N and gives higher E/N than conventional atmospheric DMS. Here we introduce the principle of vacuum DMS and show the examples of analyses using LC-vacuum DMS-MS.