Abstract

Poster Presentations

Day 3: Friday, May 20  Poster Room(Gekko)

A Study for Improving Detection Sensitivity of Volatile Compounds Using Volatimeship Combined with Direct Analysis in Real Time-Mass Spectrometry (DART-MS)

(1BioChromato, 2S&B Foods)
oChikako Takei1, Yasuyuki Semba1, Takehito Sagawa2, Yasuo Shida1

In case of volatile compounds analysis using volatimeship combined with DART-MS, we previously reported that nitrogen was used as a Volatimeship career gas and helium was used as a DART ion source career gas. In order to improve the volatile compound’s detection sensitivity, we changed Volatimeship career gas nitrogen to helium, the molecular related ions of limonene has been risen up almost doubled.