日本質量分析学会 第68回質量分析総合討論会会

演題概要

オーラルセッション

第2日 5月12日(火) 15:50~16:10 B会場(1202)

紫外線照射により劣化させたポリエチレンテレフタレートフィルムのイメージング質量分析と他の分析手法との相関

(1JEOL2産総研機能化学)
o佐藤貴弥1作田裕介1窪田梓1中村清香2Thierry, Fouqet2渡邉亮太2佐藤浩昭2上田祥久1

Matrix assisted laser desorption/ionization (MALDI) mass spectrometry is a powerful tool for the analysis of synthetic polymers. By using MALDI with a high-resolution time-of-flight mass spectrometer, this technique can be used to identify differences in polymer end groups and their molecular weight distributions. Kendrick mass defect (KMD) analysis of this data simplifies the data interpretation, even for complicated mass spectra, by showing an overview of the differences observed within a sample. MALDI mass spectrometry imaging (MSI) is a method for analyzing the spatial distribution of target compounds on a surface—typically proteins, peptides, lipids and drugs. In this report, we combined MALDI-MSI and KMD to analyze a polyethylene terephthalate (PET) film that had been degraded by ultraviolet (UV) irradiation. We also report the correlation with other analytical method such as, scanning electron microscope, pyrolysis gas chromatography mass spectrometry, infrared spectroscopy.