演題概要

オーラルセッション

第1日 5月18日(水) 16:25~16:45 C会場(星雲2)

トリプル四重極GC-MSを用いたダイオキシン類の測定

(1三浦工業2愛媛大院連農)
o山本一樹1,2稲葉健治1藤田寛之1

In the European regulation of dioxin and PCBs analysis in food and feed, use of GC-MS/MS have been permitted in addition GC-HRMS since June 2014. Accordingly, the performance and technologies of GC-MS/MS have been improving by each manufacturer. Agilent 7010 triple quadruple GC-MS was also improved due to make more high sensitivity than the previous model. We evaluated whether there is a possibility to use the system due to determine dioxins and PCBs in environmental samples.
The relative sensitivity coefficient and the detection limit of the device were calculated (from the standard deviation and S/N) with the calibration standard solutions. The results were as well as the stability and sensitivity of GC-HRMS. When it comes to the product ion scan mode with dioxin standards, Dioxin-COCl (-63)-ions were mainly detected as product ions. Furthermore, one of the advantages of GC-MS/MS measurement is that it is possible to eliminate the interferences of dioxins determination such as high level PCBs.